- Qualified electrician for electrical connection
- Plumber for drain, air and water connections
- HVAC installer (or other qualified personnel) for chamber vent connection
- Level floor
- Access to the outside via a roof vent or side wall vent
- Floor drain
- Pressurized 450 – 690 kPa (60-100 psi) ASTM D1193 reagent grade type IV water. Best chamber performance is obtained at the higher pressure ratings
- Clean dry compressed air regulated to 207 – 345 kPa (30-50 psi)
- Customer specified electric power from a lockable disconnect with appropriate amount of electric cord and conduit for the chamber location
- 1.9 cm (¾”) plastic (PVC) plumbing fittings and ball valve for chamber drain to connect the chamber to the floor drain
- 3.175 cm (1.25”)(PVC) plumbing fittings for the chamber vent
- 3.175 cm (1.25”)(copper or galvanized steel) plumbing fittings for the water jacket drain connection
- 1.27 cm (1/2”) O.D.(PVC) plumbing fittings for the chamber drain connection
- 9.525 mm (3/8”) O.D. poly tubing from the compressed air line source to the chamber compressed air installation fitting
- 9.525 mm (3/8”) O.D. poly tubing from the pressurized DI line to the chamber inlet
Improper installation will affect the operation
of the instrument and compromise test
integrity. We strongly discourage
installation in a warehouse, boiler
room, garage, factory floor, paint room,
powder mixing room, where outside doors
are frequently opened, where dust is
frequently airborne, or where humidity,
air pressure or temperature vary widely.
|